Noise & nonlinearity separation (Nonlin stage)

The Nonlin stage measures how linear your structure is. It runs a Schoukens random-phase multisine measurement: a structured set of captures that gives three things at every frequency line: the best linear approximation (BLA) of the frequency response, the level of the measurement noise, and the level of the nonlinear distortion.

An ordinary transfer function cannot give you that split. Coherence drops for noise, for a poor reference and for nonlinearity alike, with no way to tell which. The Nonlin stage answers "is my structure behaving linearly at this excitation level, and if not, where?"

What the method measures

The excitation is a multisine: a periodic signal built from many sinusoids at random phases across your band. A run plays M realisations, each with fresh random phases, and captures P repeated periods of the steady-state signal in each.

  • The scatter between periods of one realisation is measurement noise, because an identical repeated input can only give different outputs through noise.
  • The scatter between realisations is noise plus nonlinear distortion. A nonlinear system responds differently to different combinations of the same lines; a linear one adds nothing beyond the noise.

Subtracting the noise variance from the realisation variance leaves the distortion. The result is a transfer function with two extra lines, σ_NL (nonlinear distortion) and σ_n (noise), both standard deviations in the same linear units as the transfer function. See Reading the results.

The method handles one or several driven outputs against any number of measured responses, including non-square systems such as one output driving three response channels.

Run a measurement

  1. Wire each output you will drive into an input channel: a loopback cable, or a force gauge on a hammer or shaker rig. This input measures the drive.
  2. Open Nonlin. Set the band (Hz) to excite and the resolution Δf.
  3. Set the level, then M, P and transient (see Design).
  4. In excitations, tick each output to drive, and for each choose measure on ch N: the input channel wired to it.
  5. Check the run time shown beside the card's title, then press Start.
  6. The view moves to TF, showing one BLA line per excitation with dashed σ lines. Read the verdict on the Nonlin card.

Getting to the Nonlin stage

Nonlin drives an output and makes its own captures, like Acquire, so it needs a live input. It is disabled when you have only loaded a saved file. It works in the browser or served locally.

Design: setting up a run

The first row of the Nonlin card holds the design. The card starts with a band of 20 to 2000 Hz, Δf of 5 Hz, level 0.1, M 6, P 4 and transient 2. The test box sets the base name for the run's sets (bla to start with).

Band and resolution

  • band (Hz): the range to excite, from a start to an end frequency.
  • resolution / period: Δf in hertz and T in seconds are two boxes for one quantity, and you can type into either. The multisine's period is a whole number of samples, N. Typing a resolution gives N = round(fs / Δf); typing a period gives N = round(T · fs). The card shows the result beneath (N = 4096 samples · 993 lines). A typed period is rounded to whole samples, so the Δf box comes back showing the resolution you can actually have.

A finer Δf needs a longer period, and every realisation captures several whole periods, so halving Δf doubles the run. Choose the coarsest Δf that still resolves what you care about, and watch the total time.

Level

level is the RMS amplitude of each excitation. It is in volts on an NI output (up to the output rail), and a fraction of full scale (rail ±1) on a soundcard or in the browser.

A random-phase multisine's peak is several times its RMS and varies from one phase draw to the next. So Start generates every one of the M × n_exc waveforms first and checks each peak against the output rail before playing anything. If any would clip, the run refuses and asks you to lower the level.

Averaging: M, P, transient

  • M (realisations, at least 2): raise it for a cleaner split of σ_NL and σ_n. The BLA itself also improves as 1/√M.
  • P (steady-state periods per capture, at least 2): raise it for a lower noise floor σ_n, at the cost of longer captures.
  • transient: periods played and discarded before the analysis window starts, so the structure settles after each new phase draw. Raise it for heavily damped or low-frequency structures. In the browser, keep it at 2 or more (see below).

Total time

The headline beside the card's title is the run's whole duration, updated as you edit:

≈ 12.4 s
12 captures × 1.03 s

In the responses · run length group, the readout 6 realisations × 2 excitations × 6 periods (2 transient + 4 steady) shows where it comes from, so you can see which factor to cut.

The run sets its own capture length, one whole window of transient + P periods per capture, and says so on the card. It overrides the duration on Acquire while the run is going and restores it afterwards.

Excitations and responses

The excitations table has a row for each output channel the device can drive, up to eight, labelled ao0, ao1 and so on. For each row:

  • tick it to drive that output during the run; and
  • choose its x source, where the analysis reads that excitation's actual drive.

The rows you tick must start at ao0 with no gaps. Every input channel that is not carrying a measured drive becomes a response. The responses: readout lists them, so you don't assign them yourself.

Measuring the drive

Every driven output needs its drive measured on an input channel. Choose measure on ch N and wire that output into input N. This works on every device because the drive and the responses share one ADC clock: the unknown moment at which the output starts rotates all their spectra by the same phase, and that cancels in the analysis.

A commanded drive option is listed, but it is disabled on every device. The output starts at an arbitrary point relative to the capture each time, and without measuring the drive that random offset would show up as extra scatter, indistinguishable from nonlinear distortion.

What the run sets for you

A Nonlin run is stricter than an ordinary log. Start checks these before it proceeds:

  • Output rate equals sample rate. The multisine is an exact whole number of samples per period, so a different drive rate would break it. When served locally, a device that limits its output rate below your sample rate blocks the run: lower the sample rate.
  • Digital low-pass off. That mode resamples the capture, which makes the period a non-integer number of samples. The leakage would look like distortion.
  • Pretrigger disarmed. A run is a fixed-length free-running window, so pretrigger is switched off for it, a note says so, and your setting is restored afterwards.

If Start is disabled, the reason appears beside the control it concerns: for example a band above the Nyquist frequency, the same input channel used as the drive for two excitations, or no response channels left.

Browser: output latency

In the browser there is no hardware-synchronised start. Scheduling the output and the browser's own output latency eat into the window that the run treats as transient. Keep transient at 2 or more so that the discarded window clears both the structure's settling and the browser's timing slop.

Running the measurement

Start runs M × n_exc ordinary one-shot captures in sequence.

Watching it run

Progress is a grid with a row for each realisation and a cell for each excitation. Cells are outlines until they run, fill while their capture is in flight, and go solid when it lands. Beside it, capture 3/12 · ~9 s left counts captures.

stop after this capture lets the capture in flight finish, since a half-played multisine is useless, and stops before the next. Whatever landed is kept, and the grid stays so you can see what you have.

The captures land as ordinary time sets named <test> r<m>e<e> (bla r1e1), hidden in the tray and legend because there can be many. show raw captures reveals them, and hide raw captures hides them again. They are normal sets you can inspect, plot or delete.

When the captures are done, the analysis runs (computing BLA…) and the view moves to TF with one BLA line per excitation. Each is named after the response channel it came from, such as resp ch 1.

Running again: replace, or keep both

Once a run has landed, a previous run choice appears before Start.

  • replace previous (the default) removes the last run's raw captures and BLA sets when the new run starts, so iterating on a design doesn't pile up sets. A message offers Undo to bring them all back.
  • keep both leaves the last run alone and names the new one apart: bla#2 r1e1, bla#2 BLA q1 (via ch0), then bla#3, and so on. Use it for a level sweep. Changing the test name between runs has the same effect.

new run forgets the run state without deleting anything. The landed sets stay in the tray, and the raw captures are shown again.

Reading the results

A BLA result is an ordinary transfer function, so everything that works on one works on it: Bode, Nyquist, phase, real and imaginary, figure export and modal fitting.

The σ overlay

On the magnitude view (and the magnitude pane of Bode), each BLA line has two thin dashed lines on the same dB axis:

  • σ_NL, in the line's colour but dimmer: the nonlinear-distortion level.
  • σ_n, in grey: the measurement-noise level.

The σ lines switch, on the TF card and on the Nonlin card, hides both. They have no legend entry; the Nonlin card shows a small key beside the switch.

A gap in σ_NL does not mean "no distortion". It means the distortion was too small for this run to resolve at that line. More realisations, more periods or a higher level can reveal distortion that a coarser run missed.

Both σ values are per realisation: the distortion or noise level in one realisation, not the uncertainty of the plotted line. That is √M smaller, σ_BLA = σ_tot / √M. So σ shows how nonlinear the system is at this level, and you divide by √M yourself to get the error bar on the BLA.

Verdict lines

Each excitation gets a plain-English verdict, for example "linear below 800 Hz; nonlinearity dominates 800–5000 Hz — level-dependent, repeat at 2–3 amplitudes". It summarises which parts of the band have more distortion than noise. With several responses it reports the worst one.

Level sweeps

Nonlinearity depends on level, so one run tells you about one amplitude. Repeat the run at two or three levels, with keep both (or a different test name) so the runs coexist. Their sets then overlay on the TF view through the tray and legend, showing where the distortion grows with level. Each level is a separate run; there is no automatic sweep.

Saving and re-analysing in Python

Save Dataset stores everything the run produced: the raw captures and the BLA sets. Each BLA set also carries the run specification it was computed from in its bla attribute, which is the run_spec argument of calculate_bla. It holds the multisine settings (n_samples, k1, k2, p_periods, t_periods, seed, amp_rms, n_exc, M), the x mode, the input channels that carry the drive and the response channels, and the sample rate. You can recompute a saved run without re-entering the design:

import pydvma as dvma

data = dvma.load_data(filename='bench_run.dvma')
raw = [d for d in data.time_data_list if d.test_name.startswith('bla r')]
run_spec = data.tf_data_list[0].bla     # saved with the result
tf_list = dvma.calculate_bla(raw, run_spec)

raw must be in run order: realisation, then excitation. Runs kept with keep both have names such as bla#2 r1e1, so filter on that prefix for them.

You can also run a measurement without the app. Build a run_spec, then for each realisation and excitation generate the waveform with dvma.multisine_generator(settings, spec) and record it with dvma.log_data(settings, output=y). Pass the list of captures to calculate_bla. See the Acquisition and Analysis references.

Next: Modal fitting, or saving and exporting.