National Instruments Hardware

A browser can't reach NI hardware. The web logger drives it through pydvma running on the lab PC, where pydvma-serve (or dvma.launch) serves the app and talks to the NI-DAQmx driver for it. The same recording code runs when you record from Python, so this page applies to both.

Getting started

  1. Install the NI-DAQmx driver and pydvma's NI support. NI-DAQmx runs on Windows and Linux only. See Installation.
  2. Start the app on the lab PC with pydvma-serve --open, or dvma.launch() from a notebook. See Running locally.
  3. In Setup, pick your NI device from input device. NI devices are listed with any soundcards. Press Full ▾ to reveal the NI-DAQ section, described below.

pydvma-serve --list-devices shows the NI devices pydvma can see, and pydvma-serve --driver nidaq --open makes an NI device the app's Default. From Python, dvma.list_available_devices() lists the devices, and dvma.suggest_ni_settings(index) returns safe settings for one.

Devices and how they differ

Device Sampling Analogue output IEPE Notes
USB-6003 multiplexed software-timed, up to about 5 kS/s no Low cost. One converter scans the channels. No anti-alias filter.
USB-6212 multiplexed hardware-timed no One converter scans the channels. No anti-alias filter.
cDAQ-9174 chassis simultaneous hardware-timed yes, on the 9234 Delta-sigma (DSA) modules with a converter for each channel. Examples: the NI 9234 (4 inputs, IEPE, pseudo-differential only, AC coupled with a high-pass near 0.5 Hz) and the NI 9260 (2 BNC outputs, up to ±4.24 V).

Multiplexed or simultaneous matters for phase-sensitive work such as transfer functions and mode shapes. On the multiplexed USB devices the channels are sampled one after another, which leaves a small fixed time offset between them, and the maximum sample rate is shared between the channels you record. A DSA module samples every channel at the same instant. It also filters against aliasing in the converter itself, at a setting tied to the sample rate, while the USB-6003 and USB-6212 have no such filter. The digital low-pass gives them one.

A cDAQ chassis appears as a single device. Your channel count runs across its input modules in slot order, and output-only modules are skipped, so 8 channels on a chassis with two 4-channel input modules spans both. The Python guide has the channel table.

Sample rate ladders and coercion

Measurement hardware often runs only a fixed set of sample rates. A DSA module such as the 9234 runs at 51.2 kHz divided by a whole number from 1 to 31, so you can ask for a rate it doesn't have. The USB-6003 and USB-6212 divide their clocks finely enough to count as continuous, but still round a little: the 6003 runs a 48 kHz request at 48019.2 Hz.

A DSA module moves an off-ladder request to the nearest rate it has. On a 9234, 8000 Hz runs at 8533.33 Hz and 5000 Hz at 5120 Hz. pydvma reads back the rate the hardware is really running and uses it for every time and frequency axis, so your data is scaled correctly. Setup and Acquire show a note such as device runs at 8533.3 Hz (requested 8000).

The sample rate field in Setup takes any rate you type. The arrow beside it lists common rates within the device's limits; on a DSA module not every one is a rate it runs.

The output side works the same way. A DSA output module such as the 9260 also rounds its rate onto its own ladder, so pydvma resamples your stimulus onto the rate the module really runs. A 30 s sweep therefore still lasts 30 s.

Tip

Pick a rate the module runs and nothing is rounded. If you need an exact arbitrary rate, a USB-6212 or USB-6003 gets closest.

A soundcard is handled the other way round: pydvma captures at a rate the hardware runs and resamples to yours. See Capture rate and delivered rate.

NI-DAQ controls in Setup

Setup shows these in its NI-DAQ section, under Full ▾, whenever NI hardware is available.

  • terminal configuration offers default, RSE, NRSE and diff. default is RSE. If the device can't do the mode you choose, pydvma uses one it can and prints a note in the terminal where pydvma-serve is running; the app does not show it. A 9234 supports only pseudo-differential, which pydvma then uses, so leave it at default.
  • NI voltage range (±V) sets the in range and the out range, both 5 V to start with. Each is limited to the device's real range, shown beside the fields as rail in ±… out ±… V. The 9260 tops out at ±4.24 V, below the 5 V start, so its output range is set to that, with a note.
  • IEPE off or IEPE 2 mA, described below.

The input range is the full scale of the input, and a smaller range gives finer resolution. The hardware only has certain ranges, though, and rounds your request up to one it has: a 9234 is fixed at ±5 V whatever you enter, and a USB-6212 asked for ±1 V runs at ±2 V. pydvma reads back the range the device is really using, and the level meters and the clip warning judge against it.

IEPE excitation

A DSA module with built-in excitation, the 9234, can power IEPE/ICP accelerometers directly. Choose IEPE 2 mA for the sensors, or IEPE off. Those are the only two currents a 9234 accepts. Switching it on puts the inputs into AC coupling and adds about 2 seconds to the first capture while the sensors' DC bias settles.

The app's IEPE switch applies to every channel

IEPE 2 mA powers all the channels you record. Forcing excitation current into a channel that has no IEPE/ICP sensor, such as a charge amplifier's output, a signal generator or a loopback from an output, can damage what is connected. The app has no per-channel switch, and a list given to dvma.launch or pydvma-serve --settings is cut to its first entry, which then applies to every channel.

If some channels carry IEPE sensors and others don't, record from Python with one current for each channel, here IEPE on channels 0 and 1 only:

settings = dvma.MySettings(
    device_driver='nidaq', channels=3,
    iepe_excit_current_A=[0.002, 0.002, 0.0],
)

Then save the result with dvma.save_data and open the file in the app. The worked example goes through a complete recording with calibration.

Output and trigger

An output stimulus plays from the device's analogue outputs during the capture, and is limited to the output range above. A pretrigger catches transients. See Output stimulus and Pretrigger.

If a capture has gaps

If the computer falls behind the hardware, the app pins a capture integrity message after the capture. The data has gaps, and transfer functions and coherence computed from it can't be trusted, so record it again. See Capture integrity.