Once you have data, recorded or loaded from a file, the web logger has four analysis stages: Time, Frequency, TF and Sonogram. Each has a small control card above a shared, interactive plot. The plot's tray, legend, zoom toolbar and frequency navigator are described in Working with plots.

The calculations are pydvma's own analysis code, so the results match the Python calculate_* functions. They run in the browser, or in Python on your machine when the app is served locally.

Most cards start with a dataset selector: All sets or a single set. When All sets is chosen and the sets disagree on a setting, the control shows –mixed–; the first change you make applies to every set.

While a calculation runs, a computing… chip appears in the header. The first calculation of a session may show starting engine… while the analysis engine loads, which happens once.

A typical impact test

  1. On Time, check the record. If Clean Impulse is offered, press it.
  2. On TF, set in to the force channel, keep avg on within set, and press Calc TF. Check the coherence.
  3. On Fit, place the frequency window on a peak and press Fit 1. See Modal fitting.

Time

The Time stage shows the recorded time series.

  • dataset: which set Clean Impulse and Resample act on. All sets (the default) acts on every set that has time data; pick one set to act on that set only. It is the same choice as the other stages' dataset dropdowns and follows the tray.
  • input channel: the channel that Clean Impulse checks and cleans.
  • x-range: Full fits all the data; First 0.2 s shows the start of the record.
  • Clean Impulse keeps the input (force) channel up to just after the hammer pulse and fades it to zero from there, so noise or a second hit on that channel doesn't reach the transfer function. The response channels are not changed. The button appears only when the input channel looks like an impulse, meaning at most 25 % of its energy is in the second half of the record. Otherwise a note says why it is hidden. With All sets, every set whose input channel looks like an impulse is cleaned, and a note names any set left out. It is a toggle (Clean Impulse: on when applied, 1 of 2 on when only some sets are): the raw recording is kept, so pressing it again restores it. Any FFT, TF or sonogram you have already computed recomputes to match the copy that is showing. Save Dataset writes the copy that is showing.
  • Resample changes the sample rate of the set(s) in dataset. Choose another set to match (the list shows each set's rate) or type a custom rate in Hz, then press Resample. With All sets, every set not already at that rate is resampled, and one Undo reverts them all. Resampling turns Clean Impulse off for that set. Going down uses the same anti-alias filter as the logger's capture-rate conversion: 96 dB stopband at the new Nyquist frequency, zero phase. Going up uses band-limited interpolation, which adds no content above the original band. Results you have computed recompute at the new rate, and the message that follows offers Undo.

Frequency

The Frequency stage computes spectra.

  • quantity: FFT, Power, PSD or CSD.
  • window: hann (the default), hamming, flattop or none.
  • averaging: shown for Power, PSD and CSD (one FFT is not averaged). This is the resolution control.
  • CSD pair: for CSD, choose channels X and Y. A set needs at least two channels. Changing the pair redraws without recomputing.
  • Calc FFT, Calc power, Calc PSD or Calc CSD computes the result. Once a result exists it recomputes as you change settings.

Power or PSD: which to read

Power and PSD come from one calculation, shown two ways.

Power is the power spectrum. Each bin holds the mean-square amplitude in that bin, in unit², so the axis reads for example Power spectrum ((m/s²)²). Read a discrete tone here: a sine of amplitude \(A\) peaks at

\[ \frac{A^{2}}{2} \]

whatever the resolution. Broadband noise does not read correctly here, because its level grows with Δf: each bin collects a slice of the band.

PSD is the power spectral density, in unit²/Hz. It is the power spectrum divided by the analysis window's effective noise bandwidth

\[ \mathrm{ENBW} = f_s\,\frac{\sum_k w_k^{2}}{\left(\sum_k w_k\right)^{2}} \quad\text{[Hz]}, \qquad S_{xx}(f) = \frac{P_{xx}(f)}{\mathrm{ENBW}}. \]

Read a noise floor here: it does not move when you change the resolution. A discrete tone has no meaningful density, because its apparent level depends on the window.

The Live stage's PSD is a separate density, computed from the live input in its own units and not calibrated (see Live monitoring).

CSD plots the cross-spectrum magnitude |S_xy| of the chosen pair, using the convention S_xy = E[X*·Y]. It carries unit_i · unit_j, so calibration is applied to it. If no auto-power is available the plot falls back to the coherence, which is dimensionless and never calibrated.

TF: transfer functions

The TF stage estimates frequency response functions.

  • in: the input (reference) channel. Each other channel becomes an out/in line: output divided by input. The estimator is fixed; there is no H1/H2 choice.
  • window: none, hann (the default) or hamming.
  • avg: none, within set (the default: average frames of one recording) or across sets (average several recordings).
  • averaging — live: the resolution control, active when avg is within set.
  • coherence: overlays the coherence function (on by default).
  • plot type: Mag (dB), Phase, Bode (magnitude over phase), Real, Imag or Nyquist.
  • Calc TF computes the result. Once a TF exists it re-estimates as you change settings.

across sets averages the sets whose own avg is across sets; choose it with Dataset on all to average every recording. A set on within set or none is never pulled in: it gets its own TF from the same Calc TF, so a noise recording loaded beside a set of hammer hits is not mixed into their average. The input channel, window and sample rate come from the first set with this option, and the result appears on the first set that qualifies. Sets with one channel, or a different sample rate, are left out, and a note names them.

Coherence near 1 means a clean, linear, low-noise response. Dips point to noise, nonlinearity or a poor reference, as in the Python guide. Coherence needs averaging: with avg set to none it is exactly 1 and tells you nothing. It also cannot tell noise from nonlinearity. To separate the two, and to get both as numbers, run a best-linear-approximation measurement on the Nonlin stage.

Scaling: x(iω) and Best Match

The scaling group on the TF card holds two tools.

  • x(iω)^, with a power from −2 to +2, multiplies the displayed spectrum by (iω)^p. +1 turns displacement into velocity and velocity into acceleration; −1 integrates. The axis unit follows (m, m/s, m/s²).

    This changes only what is plotted. It is applied per set, never alters the stored arrays, and is saved with the set in the .dvma file. It applies to the FFT view and to every TF plot type, not to Power, PSD or CSD. The modal fit draws its fit lines with the same power, so they overlay the data as plotted, but the fit itself uses the measured values and its own TF type: the fitted frequencies, damping and modal constants do not depend on the power you view.

  • Best match rescales every other TF onto one reference TF over the frequency window you are showing, so the sets overlay. Choose the reference in ref TF, which lists the transfer functions by the name of their data card in the tray (a card with several TF lines lists each, as card · ch_2/ch_0), then press Best match. The reference itself is not changed. Every other TF line of every set is rescaled, whether or not it is showing, including the reference card's other lines. A message such as Best match onto hammer: noise ×0.42 reports the factors it applied.

    It matches the reference as displayed, calibration included, and stores each result as that line's transfer function factor, with the reference's unit. Channel calibrations are left alone. When a line it would rescale already reads at a calibrated level, it asks first, naming the reference and the measurements whose level it would replace; otherwise it just runs. Either way the message that follows offers Undo. See Best match sets the transfer function calibration.

Nyquist and Bode plots

On a Nyquist plot the card shows fmin and fmax boxes for the shared frequency window, and the frequency navigator opens automatically. The two panes of a Bode plot share a frequency axis and each has its own y control.

Sonogram

The Sonogram stage shows how frequency content changes over time. The method switch chooses STFT (the default) or CWT.

  • dataset and channel: unlike the other cards there is no All sets option. A sonogram is one channel of one set, and the list holds only sets with time data. If there are none, Calc Sonogram is disabled with a note.
  • dynamic range: the dB span of the colour map, 30 to 120 dB. It is unused, and disabled, when the plot's colour is set to linear.
  • Calc Sonogram computes the heat map. Once a sonogram exists it recomputes as you change the channel, method or window settings.

With STFT, resolution is a slider (64 to 4096 points) plus an exact nFFT box. A longer window gives finer frequency resolution and coarser time resolution.

With CWT, a complex Morlet wavelet transform, the time and frequency resolution adapt across frequency. It separates close low-frequency modes better than any single STFT window.

  • wavelet Q (w0): a slider from 4 to 64, with a box that accepts up to 128. A higher value gives finer frequency resolution and coarser time resolution. This is the main resolution control.
  • voices / octave: how densely the log-spaced frequency grid is sampled. The default, auto, follows the wavelet Q (at least 0.6 × w0, and never below 16). Choose a number to fix it.
  • freq range — Hz: optional minimum and maximum. Either box works on its own; a blank one stays automatic. Automatic runs from three decades below the top up to 0.4 × the sample rate, so 1.2 Hz to 1.2 kHz at 3 kHz; a short record starts a little higher, so that the lowest wavelet fits in it. The range applies to the damping fit as well as the picture.

The CWT is drawn on its own log-spaced grid, so the frequency axis switches to log the first time you choose it. Change the axis with y lin / log on the plot toolbar; colour dB / lin chooses how the heat is coloured (see Working with plots). STFT and CWT each remember their own frequency axis, so a frequency range and lin/log set up for one are back when you switch to it again. The time range is shared by both.

Long records and memory

A CWT damping fit holds one complex number per frequency per time sample, so a long, full-rate record over the whole band can need more memory than the analysis engine allows. The fit then stops with a message naming the remedies, such as narrowing freq range. The limit is 0.75 GiB in the browser and 8 GiB when the app is served locally.

A sonogram or damping calculation that runs longer than about three seconds shows a progress bar with a Stop button. Stop ends the calculation and restarts the analysis engine, which takes a few seconds; press Calc Sonogram again afterwards.

Fit damping

Fit damping opens the damping panel beside the sonogram (below it on a narrow screen). Choose the method with peaks | bands.

  • peaks finds spectral peaks at the fit's start time, fits each peak's free decay, and estimates a damping value Qn for each. The left chart shows the spectrum at the start time with a threshold line you can drag (or type; blank is automatic): only peaks above it are fitted. The other chart shows the measured decay of each mode with its fitted line and an f Hz, Qn=… legend. Each mode is fitted only while it stands more than 10 dB above its own noise floor, so the chart's markers stop where the noise begins. A peak with fewer than three frames above that gets no fit. It works with either sonogram method, and the CWT can separate modes that the STFT merges.
  • bands filters the decay into bands (all (broadband), octave, 1/3 octave or 1/10 decade), forms each band's Schroeder energy-decay curve, and reports EDT, T20, T30 and T60 and the band's Qn. Each band's curve stops where its decay meets the noise, marked by a tick on the chart, and the decay's own energy beyond that point is added back. Integrating on to the end of the record would count the noise as decay. A greyed value is not trusted, and hovering it says why. Either the decay starts too close to the noise for that window (20 dB for EDT, 35 dB for T20, 45 dB for T30, as ISO 3382 asks), or it is shorter than the band filter's own ringing, or its fitted line doesn't follow the curve. T60 is the first trusted value of T30, T20 and EDT. A — means the decay never spanned that window, or, for T60, that none is trusted. Save CSV writes the table with each band's range above the noise, its cut time and the flags.

Both methods share start (s), a number box (blank lets pydvma infer it: from the pretrigger, or else, for bands, from the impact, 2.3 ms after the loudest sample) and a draggable start line on the sonogram. Together they set where the free decay begins. The fit updates as you change any control.

On a wide screen the panel's charts stack in a column beside the sonogram. Click a chart, or its ⤢ button, to expand it over the plot area, and click again to restore it. Each chart has a save button that saves the chart with the Export stage's figure options (format, background and width), legend included, delivered the same way as Save Figure; the energy-decay chart keeps each band's cut tick, as on screen. The bands table saves as CSV.

Resolution and averaging

The Power, PSD and CSD cards, and the TF card with within set averaging, share one resolution control. Its four numbers are linked: change one and the others follow. There is also a slider.

  • N frames: the number of averaging frames, overlapped by 50 %.
  • frame s: the length of one frame in seconds.
  • nFFT: the samples per FFT.
  • Δf (Hz): the frequency resolution.

They are related by

frame_length = duration / (N_frames * 0.5 + 0.5)
nFFT         = round(frame_length * fs)
df           = fs / nFFT

More frames means shorter frames, coarser Δf and a smoother estimate. Fewer frames means finer Δf and a noisier estimate. This is the usual Welch trade-off. The slider covers the sensible range for the set's sample rate and duration, and the boxes accept values beyond it.

Axis labels and units

Axis labels show the units of what you are plotting. On a calibrated channel they show the engineering unit, for example Amplitude (m/s²), Power spectrum ((m/s²)²) or |H| ((m/s²)/N). Power quantities carry the square of the unit, a CSD carries the product of the two channels' units, and a transfer function carries output over input.

An uncalibrated channel shows the units of the stored samples. If the capture is genuinely in volts (an NI capture, or a soundcard whose full scale is known) the labels read Amplitude (V), V², V²/Hz and V/V. If it is in fractions of full scale, the label has no unit. See What calibration changes.

From Python

Each stage has a Python equivalent, so you can repeat or script any calculation. See the Python analysis guide.

In the app In Python
Clean Impulse clean_impulse
Resample resample_to_fs in pydvma.analysis
FFT calculate_fft
Power, PSD, CSD calculate_cross_spectrum_matrix, or calculate_cross_spectra_averaged for the averaged result. The auto-power is Pxy[i, i], and dividing it by enbw_hz gives the PSD
TF calculate_tf, or calculate_tf_averaged for across sets
x(iω)^ multiply_by_power_of_iw, which changes the data in place
Best match best_match
Sonogram calculate_sonogram (STFT), calculate_cwt (CWT)
Fit damping, peaks and bands calculate_damping_from_sono and calculate_damping_from_cwt; calculate_damping_by_band

Next: Noise and nonlinearity separation, Modal fitting, or saving and exporting.