Skip to content

Data Acquisition

This guide covers recording from Python with dvma.log_data, using sound cards and NI hardware. For point-and-click recording, start the web logger from your notebook with dvma.launch(settings) (Running locally).

Set every option when you create MySettings, as in the examples below. Python basics explains why assigning to settings.fs and the like afterwards goes wrong.

Hardware Support

  • Sound cards and audio interfaces, through the sounddevice library (pip install "pydvma[soundcard]"). On macOS, pydvma also asks CoreAudio for the device's true rate ladder and pins its clock to the capture rate while the stream runs, restoring it afterwards.
  • National Instruments DAQ, through NI-DAQmx (pip install "pydvma[ni]").

Soundcard Acquisition

Basic Setup

import pydvma as dvma

settings = dvma.MySettings(
    device_driver='soundcard',
    fs=44100,            # a typical soundcard rate
    stored_time=2.0,     # seconds per recording
    channels=2,
)

Listing Available Devices

Start here, before writing any settings:

dvma.list_available_devices()

It reports one block per physical device rather than one line per enumeration slot, and for each one it says what pydvma actually knows:

  Line (U24XL with SPDIF I/O)   [ESI U24 XL]
    calibration : CHARACTERISED  full scale 1.8819 V peak, fixed gain
    hardware    : clocks 8000/16000/32000/44100/48000 Hz
    >> index 27  Windows WDM-KS       delivers 44100/48000     24-bit, refuses rates it cannot clock
       index 23  Windows WASAPI       delivers 8000/16000/32000/44100/48000
       index 10  Windows DirectSound  delivers 44100
       index 1   MME                  delivers 44100

From the command line, the same report without starting a server:

pydvma-serve --list-devices

Three things in there are worth reading carefully.

calibration: are the readings volts, or not? Channel counts and sample rates come from the driver and are equally reliable for any interface. The voltage scale does not. CHARACTERISED means the model is in pydvma's device table, so VmaxSC is derived and readings are real volts. NEEDS GAIN means the model is known but has an analogue knob no audio API can read: state it with input_gain_db. uncalibrated means the default VmaxSC = 1.0 is a placeholder, so readings are full-scale units. Fix it from the maker's spec, or measure it with verify_input_scaling against a source of known level.

hardware vs delivers: one interface, several backends. On Windows a single device is published once per host API, and they are not equivalent. The >> marks the one pydvma recommends.

The index moves. It is a position in an enumeration, not an identity, so prefer to name the device:

settings = dvma.MySettings(device='U24XL', fs=48000, channels=2)
# note: using 'Line (U24XL with SPDIF I/O)' via Windows WDM-KS
#       (24-bit, refuses rates it cannot clock); 4 backends available
#       - device index 27

device= takes a case-insensitive substring, picks the best backend for the rate you asked for, and records the name and host API so a later capture follows the hardware if the list reorders. Ask for a rate the recommended backend cannot clock and it moves to one that can, and says so:

dvma.MySettings(device='U24XL', fs=8000)
# note: ... via Windows WASAPI ... [not the default backend:
#       Windows WDM-KS cannot clock 8000 Hz] - device index 23

It refuses rather than guess if the name matches two different devices. An index still works if you want one (device_index=27), and list_available_devices(raw=True) prints the plain index-ordered list.

Name the MODEL to make settings portable between machines

An OS names the same device differently: the ESI U24 XL is U24XL with SPDIF I/O on macOS and Line (U24XL with SPDIF I/O) on Windows, and a Scarlett 2i2 is generic Analogue 1 + 2 (Focusrite USB Audio) on Windows, which does not contain the model at all. Name the model instead and it resolves on either:

dvma.MySettings(device='ESI U24 XL', fs=48000)
# note: using 'Line (U24XL with SPDIF I/O)' (matched by model, not
#       device name) via Windows WDM-KS ...

Model matching ignores case and punctuation ('esi-u24-xl' works), is tried only when the raw name matches nothing, and is reported in the note. It works for any model in pydvma._soundcard_specs. If nothing matches, the error lists the devices that are present.

To query a configured device directly:

from pydvma import streams

streams.soundcard_device_name(settings)   # e.g. 'Scarlett 2i2 4th Gen'
streams.native_input_rates(settings)      # e.g. [44100, 48000, 88200, 96000, ...]

soundcard_device_name resolves an unset device_index the same way the recorder does, so a query and the stream that follows describe the same device. native_input_rates returns an empty list where the platform cannot answer (anything but macOS and Windows), meaning "capability unknown".

check_input_settings is not a capability probe on macOS

It approves every rate CoreAudio is willing to resample to, which is all of them: a Scarlett 2i2 accepts a 3 kHz request while its hardware ladder starts at 44.1 kHz. The conversion is then silent, with as little as 12 dB of alias rejection. Ask the hardware with native_input_rates instead. pydvma does, and captures at a rate the device really runs (see Sample Rate Selection).

Recording

data = dvma.log_data(settings, test_name='recording_01')
time_data = data.time_data_list[0]       # shape (samples, channels)

log_data returns a DataSet holding one TimeData. To collect several recordings in one dataset, add each TimeData with data.add_to_dataset(new_data.time_data_list[0]); the impact test example does this in a loop.

National Instruments DAQ

Requirements

  • NI-DAQmx driver installed
  • the nidaqmx Python package (pip install "pydvma[ni]")
  • Windows or Linux (NI provides no macOS driver)

Configuration

settings = dvma.MySettings(
    device_driver='nidaq',
    device_index=0,       # from list_available_devices(); see below
    fs=10000,
    stored_time=2.0,
    channels=4,
    VmaxNI=10,            # input range, ±10 V
)

Finding your device index

device_index is an index into the NI device list as nidaqmx enumerates it, with each cDAQ chassis collapsed to a single entry. Don't guess. The plain listing shows the chassis modules and channel counts, in exactly the order device_index expects:

dvma.list_available_devices(raw=True)
# ...
# Devices available using device_driver='nidaq', by index:
# 0: cDAQ1 (cDAQ-9174, chassis) AI=4 AO=2 modules=['cDAQ1Mod1', 'cDAQ1Mod2']
# 1: Dev1 (USB-6003, device) AI=8 AO=2

Here the chassis is device_index=0 and the USB-6003 is device_index=1. (dvma.get_devices_NI() returns a flat list that names the chassis and each module separately, so its indices do not match device_index.)

dvma.suggest_ni_settings(device_index) returns safe ranges, rate and terminal mode for whatever is at that index (see below).

Terminal Configuration

Set NI_mode to one of:

NI_mode Terminal configuration
'DAQmx_Val_RSE' referenced single-ended (default)
'DAQmx_Val_Diff' differential
'DAQmx_Val_NRSE' non-referenced single-ended
'DAQmx_Val_PseudoDiff' pseudo-differential (required by DSA modules such as the 9234)
settings = dvma.MySettings(device_driver='nidaq', NI_mode='DAQmx_Val_Diff')

cDAQ chassis with multiple modules

A CompactDAQ chassis is addressed as a single device: use the one device_index for the chassis, not one per module. channels=N is then consumed across the chassis's AI modules in slot order, so a chassis with two 4-channel AI modules (e.g. two NI 9234s) gives eight channels that span both modules automatically:

settings = dvma.MySettings(
    device_driver='nidaq',
    device_index=0,        # the chassis (one logical device)
    channels=8,            # spans both AI modules
    NI_mode='DAQmx_Val_PseudoDiff',   # required by the 9234 (see below)
    VmaxNI=5,              # the 9234 is fixed at ±5 V
    fs=12800,
)

The captured array's columns follow slot order. With a chassis whose slots are Mod1 (4-ch AI), Mod2 (AO), Mod4 (4-ch AI), the AI task skips the AO-only module and maps:

Column Physical channel
0–3 Mod1/ai0–ai3
4–7 Mod4/ai0–ai3

So an accelerometer wired to the second module's ai1 is column 5 of time_data, and any per-channel setting (iepe_excit_current_A, channel_sensitivities, pretrig_channel) is indexed the same way. AO-only modules in the middle of the chassis are simply skipped when counting AI channels (and vice versa for output).

Sensible defaults: suggest_ni_settings

suggest_ni_settings(device_index) inspects the device and returns safe, in-range values (terminal configuration, full-scale ranges, sample rate) you can pass straight to MySettings:

kwargs = dvma.suggest_ni_settings(0)        # for the chassis at index 0
settings = dvma.MySettings(channels=8, **kwargs)

For the lab cDAQ (two 9234s and a 9260 AO) this gives NI_mode='DAQmx_Val_PseudoDiff', VmaxNI=5, output_VmaxNI≈4.24, and a rate on the 9234's discrete ladder.

NI 9234 / DSA module constraints

Delta-sigma (DSA) modules like the 9234 differ from the multiplexed USB-600x/621x devices, and pydvma depends on several of their quirks:

  • Pseudo-differential only. Set NI_mode='DAQmx_Val_PseudoDiff'; other terminal modes are rejected.
  • Fixed ±5 V range. A VmaxNI other than 5 is accepted by the driver but does not change the hardware range.
  • Simultaneous sampling. Every channel has its own ADC, so all channels (across modules, via the chassis timebase) are sampled at the same instant. There is no inter-channel skew, unlike the multiplexed USB DAQs.
  • Automatic anti-alias filter. It is locked to the sample rate and not configurable. AC coupling adds a high-pass at about 0.5 Hz.
  • Discrete sample-rate ladder. The driver rounds fs to the nearest rate the module can run, rather than running arbitrary rates.

Non-standard / gappy layouts

The count-based channels=N assumes each module is filled from ai0 upward. For a non-contiguous set (skip a channel, start partway into a module, mix channels across modules), give an explicit DAQmx physical-channel string instead:

settings = dvma.MySettings(
    device_driver='nidaq',
    channels=5,                                              # must match the string
    input_channels_spec='cDAQ1Mod1/ai0:3,cDAQ1Mod4/ai1',     # inputs
    output_channels_spec='cDAQ1Mod2/ao0',                    # outputs
)

When set, these override the automatically built channel strings verbatim (nidaqmx backend only).

Triggered Acquisition

Pre-trigger Recording

Useful for capturing transient events like impacts:

settings = dvma.MySettings(
    fs=10000,
    stored_time=1.0,
    channels=2,
    chunk_size=1000,          # must be at least pretrig_samples
    pretrig_samples=1000,     # samples to keep before the trigger
    pretrig_threshold=0.5,    # trigger level (see the units below)
    pretrig_channel=0,        # channel to monitor
    pretrig_timeout=20,       # seconds to wait FOR THE TRIGGER
)

When recording starts, pydvma continuously buffers data. When the signal exceeds the threshold, it saves the pre-trigger samples plus the post-trigger duration. The first sample above the threshold lands at exactly index pretrig_samples of the returned capture.

pretrig_timeout bounds the wait for the trigger event only. Once the signal crosses, the post-trigger data is given stored_time + 5 seconds of its own to arrive, so a capture longer than the timeout is never cut short. If nothing crosses in time, log_data does not raise. It returns the most recent stored_time * fs samples, exactly as an untriggered log would.

Two constraints on pretrig_samples: it must not exceed chunk_size (that is all the pre-trigger context the buffer keeps), and it must be less than stored_time * fs (or there is no post-trigger data left to record). Both raise a ValueError naming the offending pair, when you create MySettings or when you call log_data.

pretrig_threshold is a magnitude in the units the recorder stores. On NI that is volts. On a soundcard it is volts once VmaxSC is set, and full-scale units while it is left at its default of 1.0. The default, None, means 5 % of the input's full scale, rounded to two significant figures: 0.25 V on a ±5 V NI range, 0.05 on an uncalibrated soundcard. It is the web app's default too, and it is worked out when the capture runs, so it follows an NI range the hardware rounded up. If 5 % sits close to your noise floor, set a sensible fraction of the signal you expect.

Capture integrity

Long captures are as safe as short ones: both recorders keep their buffers as circular rings, so the work per chunk is fixed however long stored_time is. If the host nevertheless falls behind the hardware (PortAudio flags dropped input, or a DAQmx task overwrites unread samples once its input buffer overflows), the loss is counted rather than hidden. acquisition.LAST_CAPTURE_OVERFLOWS holds the number of events in the last capture, log_data prints a warning, and the web logger shows a "capture integrity" message. A capture with a non-zero count has gaps, and transfer functions or coherence computed from it are not trustworthy, so repeat it.

Not every gap is flagged by the host. A USB audio driver that loses a USB packet zero-fills it, and PortAudio never sees an overflow: the capture simply contains stretches of exact digital silence (measured on a Scarlett 2i2 at anything from 8 samples to 188 ms). So log_data also scans every capture for runs of at least acquisition.DROPOUT_MIN_RUN frames in which every channel is exactly zero (acquisition.exact_zero_dropouts). A live analogue input never produces that, because its noise floor keeps the converter busy. The result is left in acquisition.LAST_CAPTURE_DROPOUTS as (count, seconds), with a warning. Leading zeros are not counted (they are the fresh-stream startup shortfall, handled separately), nor is an effectively silent record, where a 16-bit host legitimately delivers zeros. Dropouts of this kind point at the USB link: try another port or cable, without a hub.

A busy computer is the usual cause. A USB audio interface streams without retries and with a small driver buffer, so a computer that is short of memory can lose packets (the zero-filled dropouts above) or garble samples on every channel at once. That shows up as transfer function coherence collapsing on the weaker channel, often for only a few seconds. An NI device over USB or PCI does not garble samples, because its driver retries and buffers seconds of data. It can still lose data when the host falls behind, on long, high-rate or decimated captures, and then the loss is a gap, counted as above. In both cases, close other applications, keep several gigabytes of memory free, restart the kernel between batches, and shorten long high-rate captures before suspecting the interface or the cables.

Output Generation

Generate signals during acquisition, for example for transfer function measurements. The built-in generator makes 'gaussian' noise, 'uniform' noise or a 'sweep', and returns (t, output) where output has shape (samples, settings.output_channels).

settings = dvma.MySettings(
    fs=10000,
    stored_time=1.0,
    channels=2,
    output_channels=1,     # one output channel
)

amplitude is in volts, and means different things for each signal:

sig amplitude is
'gaussian' the standard deviation (rms) of the noise
'uniform' the peak, so values lie in ±amplitude
'sweep' the peak of the sine

With f=[f1, f2], noise is band-pass filtered between f1 and f2, and amplitude is then the rms of the filtered result. The waveform fades in and out over its first and last tenth (at most 0.1 s).

The output can never exceed ±settings.output_vmax() (output_VmaxNI on NI, output_VmaxSC on a soundcard). If the waveform would, the generator scales the whole waveform down to fit, so its rms drops, rather than raising an error. Check np.abs(output).max() if the level matters.

Gaussian White Noise Output

# about 0.1 V rms white noise
t, output = dvma.signal_generator(
    settings,
    sig='gaussian',
    T=settings.stored_time,
    amplitude=0.1,      # volts, rms
)

# Record with the output playing
data = dvma.log_data(settings, output=output)

Sine Sweep (Chirp) Output

# a 0.5 V peak sine sweep from 10 Hz to 1000 Hz
t, output = dvma.signal_generator(
    settings,
    sig='sweep',
    T=settings.stored_time,
    amplitude=0.5,      # volts, peak
    f=[10, 1000],       # start and end frequencies, Hz
)

data = dvma.log_data(settings, output=output)

Custom NumPy output

signal_generator is convenient but limited to three shapes, a single amplitude and an optional band. For anything else (a multi-tone, a measured or imported waveform, different drives on each channel, an MLS sequence, a stepped sine) build the array yourself and pass it to log_data(..., output=...). The format is small but strict:

Requirement Detail
Shape 2-D (N_samples, output_channels): one column per AO channel, even for a single channel (use arr[:, None]). The column count must equal settings.output_channels.
Units Volts. There is no ±1 normalisation: a value of 2.5 means 2.5 V at the terminal.
Sample rate The array is clocked out at settings.output_fs (default fs). Build the time base with 1 / settings.output_fs, and make it about stored_time long to span the capture. A sound card has one clock for input and output, so when the stimulus plays out of the device you are capturing on, log_data resamples the array onto the capture rate and rewrites output_fs to match (streams.output_shares_input_clock). The physical signal is preserved (a sweep sweeps the frequencies it was generated for) but the sample grid you built is not the one that plays. Separate devices, and NI, keep independent clocks.
Range Every sample must lie within ±settings.output_vmax(). On NI, out-of-range samples are rejected by DAQmx (error -200077).
dtype Any float. It is cast internally (to volts on NI, to ±1 float32 on the soundcard).

A hand-built array gets no ramp and no safety clamp

signal_generator fades its waveform in and out and scales it to fit the full scale for you. A raw array gets neither: you own both. A discontinuity at the first or last sample will click and can ring the structure, so window the ends yourself for transient-sensitive work, and keep the signal inside ±output_vmax().

import numpy as np

fs = settings.output_fs         # output clock (defaults to settings.fs)
T = settings.stored_time        # match the capture length
vmax = settings.output_vmax()   # full-scale output, in volts
t = np.arange(0, T, 1 / fs)

# any waveform you like, in volts: here a multi-tone of 100, 220 and
# 505 Hz, 0.3 V peak each
tones = np.array([100.0, 220.0, 505.0])
y = 0.3 * np.sin(2 * np.pi * np.outer(t, tones)).sum(axis=1)

# raised-cosine fade over the first and last 10 ms, to avoid a click
n_ramp = int(0.01 * fs)
ramp = 0.5 * (1 - np.cos(np.linspace(0, np.pi, n_ramp)))
y[:n_ramp] *= ramp
y[-n_ramp:] *= ramp[::-1]

# stay inside the rails: nothing clamps a custom array
y = np.clip(y, -vmax, vmax)

output = y[:, None]             # (N, 1): a single output channel
data = dvma.log_data(settings, output=output)

Multiple output channels. Use one column per channel, with output_channels set to match. For example a 50 Hz sine on the first output and an independent noise drive on the second:

settings2 = dvma.MySettings(fs=10000, stored_time=1.0, channels=2, output_channels=2)

a = 0.5 * np.sin(2 * np.pi * 50 * t)
b = np.clip(0.1 * np.random.randn(t.size), -vmax, vmax)
output = np.column_stack([a, b])      # (N, 2): columns map to ao0, ao1
data = dvma.log_data(settings2, output=output)

Record the drive as a reference channel

The drive is not recorded unless you wire it back into an input. Set use_output_as_ch0=True in MySettings and the played output is prepended as channel 0 of the returned data. That suits transfer functions, where you want the excitation captured alongside the response rather than assumed. The prepended column passes through uncalibrated (cal factor 1).

Volts and Full Scale

Recorded data and generated output are in volts wherever pydvma knows the hardware's scale. Time series, FFTs, transfer functions and output signals are all in volts, and become engineering units when channel_cal_factors is applied for display.

NI inputs and outputs

  • settings.VmaxNI (default 5 V) is the input range limit. NI devices are read directly in volts, so it does not scale the data. It sets the limits of the AI task (min_val=-VmaxNI, max_val=+VmaxNI), and a signal beyond it clips. Pick the smallest range that covers your signal, since smaller ranges give better resolution. DAQmx rounds your request up to a range the hardware really has (a 9234 is fixed at ±5 V and accepts any value; a 6212 asked for ±1 V runs at ±2 V). pydvma reads back the range in use, prints a note if it differs, and stores it in settings.VmaxNI.
  • settings.output_VmaxNI (default VmaxNI) is the AO task's full scale. An NI 9260 is limited to ±4.24 V, and asking DAQmx for more fails with error -200077. A signal_generator waveform above output_VmaxNI is scaled down to fit (see Output Generation). suggest_ni_settings(device_index) returns safe defaults for a device.

Soundcard inputs and outputs

sounddevice delivers samples as ±1 normalised float32. pydvma scales those to volts with a calibration constant, so downstream code sees voltages:

  • settings.VmaxSC (default 1.0) is the input full-scale voltage: the voltage at the jack that reads a normalised 1.0. The default treats normalised samples as volts at unit scale, which is no calibration. Set it to the measured full-scale voltage and acquisitions become calibrated.
  • settings.output_VmaxSC (default VmaxSC) is the output full-scale voltage: output_signal divides the requested waveform by it to recover the ±1 that sounddevice expects. Because it follows VmaxSC, a derived input full scale (below) moves the output scaling with it. On interfaces whose output level is set by an analogue knob (the Scarlett's front-panel Output control), output voltage is only repeatable at a marked knob position.

Deriving VmaxSC from the preamp gain

On a characterised interface you do not have to measure the full scale. It follows in closed form from the interface's published maximum input level L (in dBu at the minimum gain Gmin), the preamp gain G you have set and, where the interface has one, the pad P (in dB, 0 when it is out):

V_fullscale_peak = sqrt(2) * 0.7746 * 10 ** ((L - (G - Gmin) + P) / 20)

Gmin is 0 on most interfaces, which makes this (L - G). The Audient iD44's line input is the exception: its gain runs from −10 dB.

pydvma cannot read G, because it is a front-panel control no audio API exposes. State it, and VmaxSC is derived for you:

settings = dvma.MySettings(
    device_driver='soundcard',
    input_gain_db=9,        # what the front panel / Focusrite Control says
    input_mode='line',      # 'line', 'inst' or 'mic'
)

This applies to the default input device. Add device='Scarlett 2i2' to name the interface. On a Scarlett 2i2 4th Gen L is 22 dBu on line, 12 on inst and 16 on mic, and the derivation was confirmed against hardware to 0.10 dB, so no calibration run is needed. A stated gain takes precedence over an explicit VmaxSC. It applies only to interfaces characterised in pydvma._soundcard_specs, and any other device keeps the VmaxSC you gave it. Changing the gain on the hardware invalidates the calibration, so state it again when you do.

An interface with a preamp on every channel (the Audient iD44) takes one value per channel: each of input_gain_db, input_mode, input_pad and VmaxSC can be a list, and a single value still means every channel. settings.input_vmax_channels() gives each channel's full scale, and settings.channels_in_volts() says which channels are in volts. A gain may go up to 2 dB past the published range, so a unit's measured knob stop can be stated. See Acoustic measurements.

settings = dvma.MySettings(
    device='iD44', channels=4,
    input_mode='mic',                        # the default is 'line'
    input_gain_db=[60, 60, None, None],      # None: that channel in full-scale units
    input_pad=[False, True, False, False],
)

Fixed-gain interfaces: nothing to state

A characterised interface with no analogue gain anywhere in its input path has a constant full scale, so VmaxSC is derived automatically with no stated gain at all. The ESI U24 XL is the first: the default settings come out calibrated in volts (+4.7 dBu, which is 1.88 V peak, confirmed against hardware to 0.07 dB). An explicit VmaxSC still wins, since that is your own calibration.

On macOS the capture stream also pins two settings such a device does have, restoring them when the stream closes. The class-compliant input volume control is set to 0 dB (on the U24 XL it is a purely digital gain, so any other value silently rescales the data), and a capture format parked at 16-bit is raised to 24-bit (macOS defaults some interfaces to 16 and resets the choice on every sample rate change).

Not every channel a soundcard reports is an input

A Focusrite Scarlett 2i2 4th Gen advertises four inputs, but only 1 and 2 are the analogue Mic/Line/Inst inputs: 3 and 4 are a digital loopback of its own output mix. Recorded unknowingly they look like a plausible pair of channels wired to nothing. pydvma._soundcard_specs.channel_roles(name, channels) reports the role of each input for a characterised device, and the web logger warns as soon as the channel count reaches one.

IEPE / ICP excitation (NI DSA modules)

The NI 9234 (and other DSA modules with built-in excitation) can power IEPE/ICP accelerometers directly. Set the excitation current for each channel:

settings = dvma.MySettings(
    device_driver='nidaq',
    channels=4,
    iepe_excit_current_A=[0.002, 0.002, 0.0, 0.0],   # 2 mA on channels 0 and 1
)

A channel with a current above 0 is switched to AC coupling. The recorder waits about 2 s after the task starts, so the sensor's DC bias can settle through the AC-coupling high-pass before it reads. Later log_data calls with matching hardware settings reuse the live task and skip the wait.

On a multi-module chassis the list is indexed in the same slot order as the captured columns (see cDAQ chassis with multiple modules), and each current is checked against the module that actually supplies that channel. An accelerometer on the second AI module is enabled by setting the current at its column index, e.g. iepe_excit_current_A[5] = 0.002 for Mod4/ai1 in the table above.

IEPE must-knows

  • Only enable excitation on channels with an actual ICP/IEPE sensor. A charge or voltage input (a force hammer, a signal generator, a loopback from an AO output) must stay at 0.0: forcing 2 mA into a non-ICP input can damage it, and into an AO terminal it is driven back into the output.
  • The legal currents on the 9234 are exactly 0.0 or 0.002 A (off or 2 mA). Any other value raises a clear error.
  • The list is positional, one entry per channel in captured column (slot) order. A scalar applies to every channel.
  • iepe_excit_current_A > 0 needs device_driver='nidaq' and a DSA module. Soundcard inputs have no configurable excitation.
  • The web logger's IEPE switch applies to every channel. Python's per-channel list is the way to mix ICP and non-ICP channels. See IEPE excitation.

Worked example: IEPE accelerometers on a cDAQ

An end-to-end recipe for the most common DSA setup: ICP/IEPE accelerometers powered straight from an NI 9234 in a cDAQ chassis, with per-channel calibration so results come out in engineering units. The chassis is at device_index=0, its first module (cDAQ1Mod1) is a 4-channel 9234, and you have two 100 mV/g accelerometers on ai0 and ai1 plus a 2.3 mV/N force hammer on ai2:

import pydvma as dvma

# 1. Confirm the chassis index, and get safe range, rate and mode for it.
dvma.list_available_devices(raw=True)     # the chassis is index 0
base = dvma.suggest_ni_settings(0)         # PseudoDiff, VmaxNI=5, a 9234-legal fs, ...

# 2. Three channels: IEPE on the two accelerometers only, and
#    per-channel sensitivities in volts per engineering unit.
settings = dvma.MySettings(
    channels=3,
    iepe_excit_current_A=[0.002, 0.002, 0.0],  # the hammer is not ICP
    channel_sensitivities=[0.1, 0.1, 0.0023],  # 100 mV/g, 100 mV/g, 2.3 mV/N
    stored_time=2.0,
    **base,            # device_driver='nidaq', device_index=0, NI_mode, VmaxNI, fs, ...
)

# 3. Record. log_data powers the ICP sensors, switches their channels to
#    AC coupling, and waits about 2 s for the bias to settle first.
data = dvma.log_data(settings, test_name='hammer_test_01')

# 4. Samples are stored in volts. The cal factors [10, 10, 434.8] are
#    attached, so plots, FFTs and transfer functions read in
#    engineering units.
time_data = data.time_data_list[0]
print(time_data.channel_cal_factors)        # [ 10.  10. 434.78]
time_data.units = ['g', 'g', 'N']           # optional axis labels
data.plot_time_data()

This relies on three things covered above:

  • Index by capture column, not by terminal label. channels=3 uses cDAQ1Mod1/ai0:2, so list position 0 is ai0, 1 is ai1 and 2 is ai2. The same index applies to iepe_excit_current_A, channel_sensitivities and pretrig_channel.
  • IEPE only where there is an ICP sensor. The hammer's channel stays at 0.0.
  • suggest_ni_settings covers the 9234 housekeeping (PseudoDiff, VmaxNI=5, an fs on the module's ladder). Override any of its keys by listing them after **base.

Clipping detection

log_data compares each channel's peak with 95 % of that channel's own full scale (settings.input_vmax_channels(): VmaxNI on NI, the derived or stated VmaxSC on a soundcard) and prints, for example, WARNING: Data may be clipped on input 1 (channel index 0), naming each channel that came within 5 % of its limit. The check uses the raw peak, before any digital low-pass. A channel with no known full scale (an iD44's optical inputs) is judged against the converter's own full scale.

On a soundcard the check is only as meaningful as VmaxSC. With the default 1.0 it compares against a nominal unit scale. Derive VmaxSC from a stated input_gain_db (above) and the threshold becomes the interface's real full-scale voltage, so the warning fires when the converter is genuinely near clipping.

Calibration and Scaling

Two multiplicative stages stand between the converter and a plotted engineering value. VmaxSC (or, on NI, the volts read directly) turns the converter's reading into volts, and is fixed when you log. channel_cal_factors then turns volts into engineering units at display and fit time, from the per-channel channel_sensitivities. It can change at any time afterwards, because the stored samples stay in volts. Calibration and units explains the idea and what it changes in each view. This section covers the Python side.

Sensor sensitivity

Pass per-channel sensitivity, in volts per engineering unit, to MySettings when you record. log_data stores its reciprocal as TimeData.channel_cal_factors, and plotting and modal fitting multiply by those factors, so results read in engineering units (g, m/s², N, and so on) with no scaling afterwards.

settings = dvma.MySettings(
    channels=3,
    channel_sensitivities=[0.1, 0.1, 0.0023],  # V/g, V/g, V/N
)
data = dvma.log_data(settings)
data.time_data_list[0].channel_cal_factors      # [10, 10, 434.78]

A single value applies to every channel. The default 1.0 means no calibration, and every value must be non-zero, so use 1.0, not 0.0, to leave a channel uncalibrated. Sensitivities come off the sensor's calibration sheet, usually in millivolts, so divide by 1000. See the tip on calibration sheets.

Setting or correcting calibration after logging

If you recorded without sensitivities, or entered a wrong value, set the cal factor directly on the data list. It is the reciprocal of the sensitivity (engineering units per volt), because it multiplies the stored volts: a 100 mV/g accelerometer (0.1 V/g) has a cal factor of 10.

# One channel of one set (set and channel indices are both 0-based)
data.time_data_list.set_calibration_factor(10, n_set=0, n_chan=0)

# Inspect, or set every set at once
factors = data.time_data_list.get_calibration_factors()
data.time_data_list.set_calibration_factors_all(factors)

freq_data_list and tf_data_list have the same three methods. A spectrum or transfer function copies the factors when it is computed, so after changing a time data list, recompute them (or set the factors on their lists too).

Calibrating a transfer function directly

A transfer function's factor is its own. calculate_tf starts it at the ratio of the output and input channels' factors, and you can replace it when you know the transfer function's true level but not the sensors'. For a free mass, accelerance is exactly 1/m, so a measured level gives the factor. units takes the calibrated unit, either as a string or as a (response, reference) pair:

data.tf_data_list.set_calibration_factor(83.7, n_set=0, n_chan=0, units=('m/s', 'N'))
data.tf_data_list[0].units          # ['(m/s)/N']

n_chan counts the transfer function's columns, which are the output channels in order with the input left out. The time data's factors are not changed. Saving keeps the factor, and the web logger shows it in its calibrate dialog when it opens the file. Recalculating the transfer function starts again from the channel ratio, so set the factor again afterwards.

Engineering-unit labels

TimeData.units is a list with one string per channel. It passes through to calculate_fft, calculate_cross_spectrum_matrix and calculate_sonogram, and calculate_tf builds (out_unit)/in_unit for each output channel, with a compound unit in brackets, as in (m/s2)/N:

time_data = data.time_data_list[0]
time_data.units = ['g', 'g', 'N']       # set after recording if you did not pass units

Monitoring and Visualization

Live monitoring

The web logger has a live oscilloscope and FFT of the incoming signal. Use its Live monitoring view to check levels and trigger settings before committing to a recording.

For a one-off look at the live buffer from Python, use dvma.stream_snapshot while a stream is running, for example straight after a log_data call:

from pydvma import streams

snapshot = dvma.stream_snapshot(streams.REC)     # a TimeData of the live buffer

Best Practices

Sample Rate Selection

Choose a sample rate at least 2.56 times the highest frequency you care about. That matches the digital anti-alias filter pydvma uses, whose passband ends at fs/2.56. Typical choices:

  • Audio and vibration: 10 to 50 kHz
  • Ultrasonic: 100 kHz and up (on hardware that supports it)
  • Slow processes: 1 to 10 Hz

The rate you pick is the rate the data comes back at. It is not always the rate the converter runs at. Hardware only runs the rates it has: a sound card's ladder starts at 44.1 kHz, so fs=3000 is captured at 44.1 kHz and decimated to 3 kHz behind pydvma's own anti-alias filter (analysis.resample_to_fs: passband to fs/2.56, 96 dB stopband at fs/2, zero phase). streams.select_capture_fs makes that choice and names the rule it applied, and the rate the converter really ran at comes back as lpf_capture_fs on the returned settings.

These settings control it:

Field Default What it means
fs 44100 The rate the logged data ends up at. Not necessarily a rate the hardware runs
capture_fs None Force the converter's rate (Hz). It must be at least fs, and is rounded up to a real rate where the ladder is known
lpf_on False Digital low-pass: capture above fs deliberately (anti-aliasing plus about 10·log₁₀(M) dB of broadband-noise process gain), then resample down
oversample 'auto' How far above fs to capture: 'lowest' (the first rate at or above 2.56 × fs) or 'highest' (as fast as the device goes)
lpf_capture_fs none Written onto the returned settings: the rate the converter really ran at, whenever it differed from fs

'auto' chooses on the physical fact. It picks 'lowest' where the converter anti-aliases in silicon (any audio interface, and NI DSA modules like the 9234), because content above the capture Nyquist is already gone before the ADC and capturing faster rejects nothing extra. It picks 'highest' on a filterless multiplexed device (USB-6003/6212), where a high capture rate is the only alias protection there is.

Duration Selection

The frequency spacing of a spectrum from one frame is 1/stored_time. To get a spacing df, record for at least this long:

df = 1.0                        # Hz spacing wanted
N_frames, overlap = 1, 0.5      # frames averaged in the analysis (1 = a single frame)
stored_time = (N_frames + 1) * (1 - overlap) / df       # 1 s here; 4.5 s for N_frames=8

Averaging frames makes each one shorter, so it needs a longer record for the same spacing (see Averaging for Better Estimates).

Anti-aliasing

Know which kind of front end you have. A delta-sigma converter (any audio interface, and NI DSA modules such as the 9234) anti-aliases in silicon at its own rate, so content above the capture Nyquist is gone before the ADC. streams.hardware_antialiases(settings) reports this per device. A multiplexed SAR device (NI USB-6003/6212) has no such filter, and anything above Nyquist folds into your band at sampling time, where no later filtering can separate it. There, set lpf_on=True so the capture runs fast and pydvma filters before dropping the rate, or choose a sample rate high enough that nothing real lives above fs/2.

Troubleshooting

No Signal Detected

  1. Check connections and that the sensor is powered (IEPE sensors need excitation on an NI 9234).
  2. Run dvma.list_available_devices() and check device or device_index, and that channels matches what you wired. On a Scarlett 2i2, inputs 3 and 4 are not analogue inputs.
  3. Check the input range or gain.
  4. Test with a known signal source.

Clipping or Saturation

log_data prints WARNING: Data may be clipped when a sample comes within 5 % of the input limit.

  • Reduce the signal amplitude, or the preamp gain.
  • Raise VmaxNI, or on a soundcard check VmaxSC matches the gain.
  • Check the sensor's sensitivity and range.

High Noise Floor

  • Ground properly, and shield cables for low-level signals.
  • Keep signal cables away from power cables.
  • Use differential inputs where the hardware has them.
  • Reduce the gain if you can, and check for ground loops.

Trigger Not Working

  • The trigger fires when the magnitude of the signal on pretrig_channel exceeds pretrig_threshold, in the units the recorder stores (volts once VmaxSC is set). Check the level against your signal.
  • Check pretrig_channel, and that the signal reaches it.
  • If nothing crosses within pretrig_timeout, log_data returns an untriggered capture rather than raising.
  • A ValueError about pretrig_samples means it exceeds chunk_size, or leaves no post-trigger data.

Warnings about gaps or dropouts

A capture integrity or dropout warning means samples were lost. See Capture integrity and repeat the recording.

Next Steps